New Study Shows Lactate Plus Provides Accurate and Reproducible Measurements of Blood Lactate
Posted September 6th, 2013.
Objectives:
The aims of this study were to: (1) determine the validity and reliability of the Nova Biomedical Lactate Plus portable analyzer, and quantify any fixed or proportional bias; (2) determine the effect of any bias on the determination of the lactate threshold and (3) determine the effect that blood sampling methods have on validity and reliability.
Results:
Though measurements from both instruments were correlated (r=0.91), the differences between instruments had large variability (SD=1.45 mM/l) when blood was sampled directly from finger. This variability was reduced by∼95% when both instruments measured blood collected in the capillary tubes. As the proportional and fixed bias between instruments was small, there was no difference in estimates of the lactate threshold between instruments. Reliability for the portable instrument was strong (r=0.99, p<0.05) with no proportional bias (slope=1.02) and small fixed bias (−0.19 mM/l).
Conclusions:
The Lactate Plus analyzer provides accurate and reproducible measurements of blood lactate concentration that can be used to estimate workloads corresponding to blood lactate transitions or any absolute lactate concentrations.